Welcome To Padma Scientific Co

Auto ELISA Processor BIOBASE2001

৳ 0.00

Auto ELISA Processor BIOBASE2001 is a fully automaticdetection equipment that integrates sample addition,incubation, shaking, washing, and reading. lt is widely usedin hospital laboratories, disease prevention and controlcenters, blood centers, and blood banks, animal and plantinspection and quarantine institutions, scientific researchinstitutions, university laboratories and other places.

Call for Price: +88 01720 342 211

Description

The BIOBASE 2001 is a high-throughput, fully automated immunoassay analyzer designed for large-scale diagnostic laboratories. It combines precise liquid handling, advanced liquid level and clot detection, multi-channel operation, and intelligent plate handling with biosafety features—ensuring high performance, flexibility, and safety in ELISA-based testing workflows.

Key Features

  1. Pressure-Based Liquid Detection: Utilizes air pressure sensing for real-time liquid level and clot detection.

  2. TIP and Clot Detection: Ensures accurate sampling and minimizes contamination risk.

  3. Dual Pipetting Channels: Two-channel pipetting system increases testing efficiency and throughput.

  4. High-Precision Pipetting: Supports volumes from 10 μL to 1000 μL with outstanding accuracy and repeatability.

  5. Gripper Monitoring: Real-time status detection with adjustable spacing and strength to handle various microplate sizes and covers.

  6. Dual Thread Control: Allows simultaneous operation of sampling arms and grippers without interference.

Sample & Reagent System

  • Sample Rack: 6 × 20 push-pull racks with automatic detection

  • TIP Racks: 3 racks for 1000 μL disposable tips

  • Pipetting Volume: 10 μL–1000 μL

  • Pipetting Precision:

    • 100 μL: CV ≤ 1%, Accuracy ≤ 2.5%

    • 1000 μL: CV ≤ 0.5%, Accuracy ≤ 1%

  • Reagent Positions: 4 racks with reagent tank detection

Washing Unit

  • Washing Head: 16-channel, 32-pin dual-row manifold with leakage recovery

  • Injection Uniformity: CV < 1.5%

  • Residual Volume: <1 μL

Reading Unit

  • Channels: 8 independent photometric channels

  • Absorbance Range: 0.000–4.000 Abs

  • Linear Range: 0.000–3.000 Abs

  • Repeatability: CV ≤ 1.0%

  • Stability: ≤ ±0.003 Abs

  • Reading Accuracy:

    • ≤ ±0.02 Abs (0.0–1.0 OD)

    • ≤ ±0.03 Abs (1.0–2.0 OD)

  • Sensitivity: ≥ 0.01 mg/L

  • Channel Difference: ≤ 0.02 Abs

  • Optical Filters: 4 standard filters (405, 450, 492, 630 nm); up to 10 filter positions available

  • Spectral Range: 400–700 nm

Incubation Unit

  • Incubators: 8 shaking incubators with dry heating

  • Temperature Range: RT to 60 °C

  • Temperature Uniformity: ±0.5 °C

  • Fluctuation: ±0.5 °C

  • Shaking Mode: Horizontal circular oscillation with 1–3 speed levels, 2.5 mm amplitude

  • Cover Station: Holds 8 covers

Data Management & Connectivity

  • Software Compatibility: Windows 7 or above

  • LIS System: Bi-directional with HL7 protocol support

  • Result Management: Absorbance and test results viewable via software

  • Printer: Optional external printer

  • Built-in Scanner: Optional

  • IAP Function: Firmware upgradeable online

Safety Features

  • Biosafety Protection: Optional negative pressure working zone with HEPA filtration

  • Disinfection: Integrated UV lamp

  • Lighting: Internal LED illumination

Working Conditions

  • Power Supply:

    • Standard: AC220V ±10%, 50/60Hz

    • Optional: AC110V ±10%, 60Hz

  • Operating Temperature: 10–40 °C

  • Humidity Range: 30–80%

Dimensions & Weight

  • External Size (W×D×H): 1275 × 750 × 1650 mm

  • Net Weight: 330 kg

Packaging Details:

  • Machine: 1445 × 1130 × 920 mm (220 kg)

  • Base Stand: 1445 × 1005 × 920 mm (125 kg)

  • Accessories: 1065 × 510 × 895 mm (30 kg)

Contact us for best price